Leysop 3mm RTP Pockels Cell
This oscilloscope trace shows the measurement of the optical transmission in a single-pass setup
measured with a photodiode of risetime 1nsec.
Parameters: 1064nm, 1.2kV, time difference of On/Off triggers 50nsec.
This oscilloscope trace shows the measurement of the optical transmission in a single-pass setup
measured with a photodiode of risetime 1nsec. In order to reduce rise and fall times,the time
difference between the On/Off triggers was reduced, at the same time increasing the drive voltage
to the cell.
Parameters: 1064nm, 1.5kV, time difference of On/Off triggers 5nsec.
This oscilloscope trace shows the optical measurement of a 10Mhz burst of pulses using this
3mm RTP Pockels cell in a single-pass setup measured with a photodiode of risetime 1nsec.
Note the time difference between the pulses of about 100nsec and the instantaneous start
of the pulse sequence. The expanded trace below shows an expansion
of the start phase.
Parameters: 1064nm, 1.5kV, 10MHz, Burst length 10msec